Bedside Infusion Technology : Syringe Pumps
Injectomat Agilia
Injectomat Agilia, the new generation of stand alone syringe pumps.
Features:
  Flow rate range: 0.1- 200 ml/h (depending on the syringe capacity - up to 1200 ml/h).
  Flow rate accuracy: ± 1% on mechanism ; ± 2 % on syringes.
  Syringe capacities: 5, 10, 20, 30/35, 50/60 CC.
 
Infusion Modes:
ml/h (0.1 ml/h increments).
Volume / time: Volume to infuse 0.1 to 99.9 ml / Time: 0 h 01 to 96 h 00.
Volume limit: 0.1-999.9 ml (0.1 ml increments).
  Bolus: Manual bolus: Rate 50-1200 ml/h (50 ml/h increments).
  Drug library: Initial drug list of 50 agents adjustable with Vigilant®, the IV Medication Safety Solution. Adjustment of the dilution, authorized infusion modes, default infusion modes, flow rate and bolus limits in ml/h.
  Pressure modes: 2 modes available: variable or 3 pre-set levels - From 100 to 900
mmHg, 50 mmHg increment.
  DPS: The Dynamic Pressure System - DPS - warns of pressure variations. A risk
of obstruction or a possible leak in the infusion line can thus be anticipated.
  Anti-bolus system: Reduces significantly bolus after occlusion release (0.2 ml
max).
  Dimensions (H/W/D/) / Weight: 160 x 345 x 135 mm / 2150 g.
  Battery: NI/MH / Battery capacity minimum: 10 hours at 5 ml/h / Battery replacement: separate compartment – no need for electrical security testing.
  Power supply: 100 – 240 Vac / 15 VA, external power 9 Vdc by a converter.

Injectomat MC Agilia
Features:
  Flow rate range: 0.1-1200 ml/h (depending on the syringe capacity).
  Flow rate accuracy: ± 1% on mechanism ; ± 2 % on syringes.
  Syringe capacities: 5,10, 20, 30/35, 50/60 CC.
 
Infusion Modes:
ml/h.
Dose rate:
Dose units: ng, μg, mg, cal, kcal, U, kU, mmol, mol, Kmol / kg.
Time units: 24h / h / min.
Dilution setting: X units / ml or X units / Y ml.
Volume or dose / time: 0.1-99.9 units ; 00h01- 96h00.
Volume limit: 0.1-999.9 ml.
  Bolus: Manual bolus: Rate: 50 -1200 ml/h; 50 ml/h increment. Programmed bolus (dose or volume/time): 0.1-99.9 units / 0.1-1200 ml/time auto-calculation.
  Programming modes: No drug names: infusion without any display of the drug names. Drug labeling: infusion with the drug names. Vigilant® Drug’Lib: the secured IV Administration mode with a drug library.
  Drug labeling: Display of the drug name based on a list of 50 pre-set drug names.
  Drug libraries: Drug parameters, infusion, bolus and loading dose parameters, authorized infusion modes, infusion mode per default. Inititial drug library of 50 drugs adjustable with Vigilant®, the IV Medication Safety Solution. Up to 120 drugs (depending on comment size) storable in 4 different drug libraries.
  Pressure modes: 2 modes available: variable or 3 pre-set levels - From 100 to 900
mmHg, 50 mmHg increment.
  DPS The Dynamic Pressure System - DPS - warns of pressure variations. A risk of obstruction or a possible leak in the infusion line can thus be anticipated.
  Anti-bolus system: Reduces significantly bolus after occlusion release (0.2 ml max).

Injectomat TIVA Agilia
Features:
  Flow rate range: 0.1-1200 ml/h (depending on the syringe capacity).
  Flow rate accuracy: ± 1% on mechanism ; ± 2 % on syringes.
  Syringe capacities: 5,10, 20, 30/35, 50/60 CC.
 
Infusion modes:
ml/h.
Dose rate:
Dose units: ng, μg, mg, cal, kcal, U, kU, mmol, mol, Kmol / kg.
Time units: 24h / h / min.
Dilution setting: X units / ml or X units / Y ml.
Volume or dose / time: 0.1-99.9 units ; 00h01- 96h00.
Volume limit: 0.1-999.9 ml.
  TCI Functions: Injectomat® TIVA Agilia is an IV anaesthesia system incorporating pharmacokinetics models for Propofol, Remifentanil, Sufentanil in plasma or effect mode.
  Bolus: Manual bolus: Rate: 50 -1200 ml/h; 50 ml/h increment. Programmed bolus (dose or volume/time): 0.1-99.9 units / 0.1-1200 ml/time auto-calculation.
  Programming modes: No drug names: infusion without any display of the drug names. Drug labeling: infusion with the drug names. Vigilant® Drug’ Lib: the secured IV Administration mode with a drug library.
  Drug labeling: Display of the drug name based on a list of 50 pre-set drug names.
  Drug libraries: Drug parameters, infusion, bolus and loading dose parameters, authorized infusion modes, infusion mode per default. Inititial drug library of 50 drugs adjustable with Vigilant®, the IV Medication Safety Solution. Up to 120 drugs (depending on comment size) storable in 4 different drug libraries.
  Pressure modes: 2 modes available: variable or 3 pre-set levels - From 100 to 900 mmHg, 50 mmHg increment.
  DPS The Dynamic Pressure System - DPS - warns of pressure variations. A risk of obstruction or a possible leak in the infusion line can thus be anticipated.
  Anti-bolus system: Reduces significantly bolus after occlusion release (0.2 ml max).


Master PCA
Features:
  5 PCA/PCEA modes, titration or loading dose functions.
  Programming memorization capability of up to 5 protocols.
  Historical last 5 patients’ data storage; visualization or print-out and down load to PC via RS232 line.
  Concentration 0.1 μg/ml to 99.9 mg/ml, 0.5 μg, 1 μg and 1 mg increments.
  Bolus dose «off», 1 μg to 99.9 mg, 0.5 μg, 1 μg and 0.1 mg increments.
  Bolus duration «flash», 30 seconds or from 1 to 15 minutes, 1 minute increment.
  Lockout interval 1 to 90 minutes, 1 minute increments; from 2 to 12 hours, half hour increments.
  Continuous rate «off» or from 1 μg/h to 99.9 mg/h, variable increments depending on the concentration.
  Dose limit 10 μg to 9999 mg, 10 μg and 1 mg increments.
  Programmable duration from 1 to 12 hours.
  Loading dose «off», from 1 μg to 99.9 mg, 0.5 μg, 1 μg and 1 mg increments.
  Infusion pressure adjustable up to 900 mmHg (or 1100 mmHg optional).
  Flow rate accuracy ± 1% on device/± 2% on syringe.

Pilot C
Features:
  Flow rates: 0.1 to 800 ml/h.
  Syringe selection 12 types of Luer lock 50/60 ml and 20 ml, 10 ml with accessory.
  Flow rate accuracy ± 1% on drive mechanism/± 2% on syringes.
  Name of the drug 15 drug names proposed, easy to customize.
  Bolus rate programmable from 50 up to 800 ml/h, with infused volume display, protected access.
  Two occlusion alarm programming modes: variable mode: numerically selected occlusion release (from 100 to 900 mmHg). Pre-selection mode: 3 preset pressure occlusion alarm levels (300/500/900 mmHg).
  Line disconnection alarm drop in pressure in the infusion line.
  Anti-bolus system: bolus volume reduction at occlusion release.
  Volume limit and total volume infused.
  RS232 serial link for standard computer interface. TTL link for Master extension.
  Battery life: 7 hours minimum, at 5 ml/h for a 50/60 syringe, displayed in hours/minutes
  Mains supply 12–15 V dc 15 W.
  Dimensions/Weight W=120, H=330, D=155 mm/2.2 kg.